Screen:12.1 inch color LCD screen
Throughput: 60T/hr
Linearity range :Linear correlation coefficient (r)≥0.99
Accuracy and fluctuation of temperature control in reaction zone:The temperature accuracy should be within ± 0.5℃ of the set value, and the fluctuation should not exceed 0.5℃
Repeatability of metering value:CV≤2%
Stability of metering value:The change in luminous value should not exceed ±3%
Power requirement :voltage:220V~, frequency:50Hz, Rated input power:600VA Fuse:F6.3AL250V
Operation environment:Temperature 10-30°C;Humidity:30%-85%
Dimensions:626mm (length) × 627mm (width) × 594mm (height)
Weight:70(kg)
Based on the chemiluminescence immunoassay method, the detection of different chemiluminescence items is realized. A small amount of blood sample is added during the test, and the test substance in the sample reacts with the reagent to generate a chemiluminescence signal which is detected by the photomultiplier tube in the optical module. The software system compares and converts the luminescence signal according to the standard curve to obtain the sample detect the substance concentration value.
1.Sample scanning
2.Sampling
3.Loading the sample rack
4.Click the “start”